Cromatografía - Laboratorio - Medio Ambiente desde 1976 ecosta@icon-sa.com iconargentina@gmail.com www.icon-sa.com www.iconargentinaonline.com Alquiler con opción a compra - Asesoramiento - Desarrollo - Servicio - Venta Lista de productos para el medio ambiente Lista de productos con sus vínculos Ubicación y como llegar a Icon Argentina IyE Ventas online
|
|
******************
|
For any model GC chassis except the 110, a single channel PeakSimple data system is standard equipment. If you'll be ordering more than one detector on your GC, you will need a four or six channel data system. The one and four channel data systems connect to your Windows™ computer's serial port, and the six channel model connects to the USB port.
2. Choose your detectors. Detector types are selected depending on the particular measurement application , the required detection limit, matrix interferences, and/or regulatory guidelines. Since all five of our chassis can mount up to four detectors simultaneously (some can mount five or even six), you can perform a surprising variety of applications with one instrument.
DETECTOR
SELECTIVITY
SENSITIVITY
(approximate)
CCD
hydrogen and hydrocarbons
500ppm
TCD
universal
200-500ppm
FID
hydrocarbons
1ppm
DELCD
chlorinated and
brominated molecules
10ppb
FID/DELCD
hydrocarbons, chlorinated
and brominated molecules
10ppm
HID
universal, except neon
10ppm
PID
aromatics and molecules
with double carbon bonds
10ppb
NPD
nitrogen and phosphorus
10ppm
NPD/DELCD
nitrogen, phosphorus,
chlorinated and brominated molecules
100ppb
TID
nitro functional groups
(TNT, etc.; explosives)
1ppb, 50ppb
ECD
electronegative compounds
10ppb
RGD
reducing gases (like
hydrogen and CO)
10ppb
FPD
sulfur and phosphorus
200ppb and 10ppb
FPD/FID
sulfur, phosphorus, and
hydrocarbons
200ppb, 10ppb, and 100ppm
Dual FPD
sulfur and phosphorus
simultaneously
200ppb and 10ppb
FID Dual FPD
hydrocarbons, sulfur and
phosphorus simultaneously
100ppm, 200ppb, and 10ppb
3. Choose your injectors. Injector types are selected by the user depending on the particular measurement application, detection limit, and regulatory requirements. Each GC chassis except the 110 comes standard with an On-Column injector. The Heated Flash Vaporization, Heated Split/Splitless, and PTV injectors are upgrades to the standard On-Column injector. There are 17 injector types from which to choose. Up to five injectors may be mounted simultaneously on Models 8610C and 8610D. The Model 310 will accommodate a single On-Column, Heated Flash Vaporization, Heated Split/Splitless, or PTV injector. Rack-mountable Model 410 can accommodate a gas sampling valve and a single On-Column, Heated Flash Vaporization, Heated Split/Splitless, or PTV injector.
Sample Types and Appropriate
Injectors
LIQUIDS:
On-Column, Heated Flash
Vaporization, Heated Split/Splitless, PTV, Heated Static Headspace, Purge &
Trap, Liquid Injection Valve, Liquid Autosamplers, or Headspace Autosampler
SOLIDS: Thermal Desorber, Heated Static Headspace, PTV, Method 5035 Purge & Trap, or Headspace Autosampler GASES: On-Column, Gas Sampling Valve, Method TO-14 Air Concentrator, or Heated Static Headspace SPME FIBERS: Heated Flash Vaporization with Low Volume SPME Liner, or Heated Split/Splitless 4. Accessorize.
Although multiple injectors and detectors can be built into an SRI GC chassis, there are instances where certain components would occupy the same space, or we simply run out of room. If you are not sure if everything you want will fit, call one of our knowledgeable technical support agents for help. We may be able to make your dream system, but there will be an additional custom configuration charge. |
|